How to Resolve an "Err 8.1" Reading on an OHaus Scale

What to do if your OHaus scale is reading "ERR 8.1", "ERR 8.2" or "ERR 8.4"

This may be an issue with "Capacity" being set to 0 or with incorrect units

  1. Go to "Main Menu" > "Setup"
  2. Go to "Capacity Unit"
  3. Set the Capacity Unit to pounds
  4. Go to "Capacity and Graduation" and set this to 500
  5. Set "Power On Zero" to OFF
  6. Go back to the home menu
  7. Press the wrench icon in the lower left hand corner of the screen, and set
  8. Exit the menu, and the error should be gone